Fluke 96040A Low Phase Noise Reference Source
Fluke 96040A Low Phase Noise Reference Source enables you to simplify your RF calibration system by replacing many of the instruments and accessories that make up your current system. It's precision signal level and attenuation, high signal purity, and precision low distortion modulation make this reference source superior to the general-purpose signal generators that are often used to calibrate spectrum analyzers, modulation meters, RF sensors, attenuators, and similar instruments. Its low phase noise provides superior phase noise performance.
Unlike many RF calibration solutions, the 96040A is designed specifically for RF calibration, with a calibration-oriented user interface that makes it easy to learn and use. The 96040A speeds up calibration procedures, reduces opportunities for operator errors, and greatly simplifies RF metrology. As the core of an RF and microwave calibration system, the 96040A covers the majority of test points required for calibrating spectrum analyzers of any frequency range.
Automated with MET/CAL® Plus Calibration Management Software, the 96040A reduces complexity and calibration times, dramatically improving efficiency and increasing capacity by 50 % or more over manual methods.
If you intend that you buy this product you will gain because Iconic Engineering always provides Fluke original products through Tools Bangladesh.
Specifications: Fluke 96040A Low Phase Noise Reference Source
96040A summary specifications* |
|
Frequency specifications |
Level specifications (50 Ω output, see extended specifications for 75 Ω) |
Range |
1 mHz to 4 GHz |
-130 to +24 dBm to 125 MHz, 14 dBm at 4 GHz (leveled) |
Resolution |
10 µHz |
0.001 dB |
Accuracy |
± 0.05 ppm ± 5 µHz |
Down to -48 dBm: ± 0.03 dB to 100 kHz, ± 0.05 dB to 128 MHz, ± 0.3 dB at 4 GHz 10 MHz to 128 MHz: ± 0.05 dB to -48 dBm, ± 0.1 dB to -84 dBm , ± 0.7 dB at -130 dBm |
Attenuation |
|
± 0.02 dB to 49 dB, ± 0.15 dB at 110 dB Relative to +10 dBm, 10 Hz to 128 MHz |
VSWR |
≤ 100 MHz: ≤ 1.05, ≤ 2 GHz: ≤ 1.1, 2 GHz to 4 GHz: ≤ 1.0 + 0.05 xf (GHz) |
Harmonics and spurious |
–60 dBc harmonics, –78 dBc spurious to 1 GHz |
Phase noise at 1 GHz |
–144 dBc/Hz, typical, at 10 kHz to 100 kHz offset |
Modulation |
AM, FM, PM, internal and external. Frequency pull and external leveling |
Frequency sweep |
1 mHz to 4 GHz. Linear or Logarithmic. Stop-Start or Center-Span |
Frequency counter |
Internal 50 MHz frequency counter |
Temperature |
Operating: 0 °C to 50 °C, 23 °C ± 5 °C for specified performance; Storage: -20 °C to +70 °C. |
Standard interfaces |
IEEE-488.2 (GPIB) |
GPIB command emulation |
9640A, 9640A-LPN, 9640A-LPNX, HP3335, HP8662A, HP8663A. |
Dimensions (HxWxD) |
146 mm x 442 mm x 551 mm (5.8 in x 17.4 in x 21.7 in) including handles Industry-standard 19 in (483 mm) rack mounting when fitted with Y9600 rack mounting kit |
Weight |
18 kg (40 lb) |
*Key specifications summary. Refer to the extended specifications for full and detailed specifications. |