Model: Fluke Calibration 9142 Field Metrology Well
Why you buy Fluke 9142 Field Metrology Well because it's working as Metrology performance in accuracy, stability, uniformity, and loading. Now it's available at Tools Bangladesh that is your best choice.
The 9142, 9143, 9144 Field Metrology Wells extend tall accomplish to the industrial process feel by maximizing portability, promptness, and functionality once tiny compromise to metrology feint.
Field Metrology Wells are packed when functionality and are remarkably within getting your hands on to use. They are lightweight, little, and fast to achieve temperature set points, still, they are stable, uniform, and accurate. These industrial temperature loop calibrators are strong for the stage transmitter loop calibrations, comparison calibrations, or to hand checks of thermocouple sensors.
With the processed substitute, there is no compulsion to carry expand tools into the arena. This optional built-in two-channel readout reads resistance, voltage, and 420 mA current as soon as 24-volt loop triumph. It as well as has vis--vis speaking-board automation and documentation. Combined, the three models (9142, 9143, and 9144each subsequent to a process substitute) lid the broad range of 25 C to 660 C.
This product is the lowest price at Iconic Engineering and Keep it simple and get the best product that you are looking for. Fluke brand that is the heigh class product and that are available now to Toos Bangladesh.
Specifications: Fluke Calibration 9142 Field Metrology Well
Base Unit Specifications |
Temperature range at 23°C |
–25°C to 150°C (–13°F to 302°F) |
Display accuracy |
±0.2°C full range |
Stability |
±0.01°C full range |
Axial uniformity at 40 mm (1.6 in) |
±0.05°C full range |
Radial uniformity |
±0.01°C full range |
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
±0.006°C full range |
Hysteresis |
0.025 |
Operating conditions |
0°C to 50°C, 0% to 90% RH (non-condensing) |
Environmental conditions (for all specifications except temperature range) |
13°C to 33°C |
Immersion (well) Depth |
150 mm (5.9 in) |
Insert OD |
30 mm (1.18 in) |
Heating time |
16 min: 23°C to 140°C 23 min: 23°C to 150°C 25 min: –25°C to 150°C |
Cooling time |
15 min: 23°C to –25°C 25 min: 150°C to –23°C |
Resolution |
0.01° |
Display |
LCD, °C or °F user-selectable |
Size (H x W x D) |
290 x 185 x 295 mm (11.4 x 7.3 x 11.6 in) |
Weight |
8.16 kg (18 lb) |
Power requirements |
100 V to 115 V (±10%) 50/60 Hz, 635 W 230 V (±10%) 50/60 Hz, 575 W |
Computer interface |
RS-232 interface |
Calibration |
NVLAP accredited calibration included |
Automation |
Use the 1586A Super-DAQ to automate temperature sensor calibration |
-P Specifications |
Built-in reference thermometer readout accuracy (4-Wire Reference Probe)1 |
±0.010°C at -25°C ±0.015°C at 0°C ±0.020°C at 50°C ±0.025°C at 150°C ±0.030°C at 200°C ±0.040°C at 350°C ±0.050°C at 420°C ±0.070°C at 660°C |
Reference resistance range |
0 ohms to 400 ohms |
Reference resistance accuracy2 |
0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading |
Reference characterizations |
ITS-90, CVD, IEC-751, Resistance |
Reference measurement capability |
4-wire |
Reference probe connection |
6-pin Din with Infocon Technology |
Built-in RTD thermometer readout accuracy |
NI-120: ±0.015°C at 0°C PT-100 (385): ±0.02°C at 0°C PT-100 (3926): ±0.02°C at 0°C PT-100 (JIS): ±0.02°C at 0°C |
RTD resistance range |
0 ohms to 400 ohms |
RTD resistance accuracy2 |
0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading |
RTD characterizations |
PT-100 (385),(JIS),(3926), NI-120, Resistance |
RTD measurement capability |
4-wire RTD (2-,3-wire RTD w/ Jumpers only) |
RTD connection |
4 terminal input |
Built-in TC thermometer readout accuracy |
Type J: ±0.7°C at 660°C Type K: ±0.8°C at 660°C Type T: ±0.8°C at 400°C Type E: ±0.7°C at 660°C Type R: ±1.4°C at 660°C Type S: ±1.5°C at 660°C Type M: ±1.4°C at 660°C Type L: ±0.7°C at 660°C Type U: ±0.75°C at 600°C Type N: ±0.9°C at 660°C Type C: ±1.1°C at 660°C |
TC millivolt range |
–10 mV to 75 mV |
Voltage accuracy |
0.025% of reading + 0.01 mV |
Internal cold junction compensation accuracy |
±0.35°C (ambient of 13°C to 33°C) |
TC connection |
Small connectors |
Built-in mA readout accuracy |
0.02% of reading + 2 mV |
mA range |
Cal 4-22 mA, Spec 4-24 mA |
mA connection |
2 terminal input |
Loop power function |
24 V DC loop power |
Built-in electronics temperature coefficient (0°C to 13°C, 33°C to 50°C) |
±0.005% of range per °C |
1. The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors. 2. Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. |